Agilent Technologies AGILENT U2761A User Manual

Page 110

Advertising
background image

92

U2761A User’s Guide

2

Features and Functions

1 Select the Arbitrary waveform function.

Click

Arb

to select the Arbitrary function.

2 Set the waveform period.

Select

Period

and use the numeric keypad or

numeric drop down

to set the period for the

waveform.

For example, set the period of the waveform to 10 ms.

3 Set the waveform voltage limits.

Select

HiLevel

and

Lolevel

to set the

high and low voltage levels that can be reached while
editing the waveform. The upper limit must be greater
than the lower limit.

For example, set the upper limit to 2.0 V and the lower
limit to 0 V.

4 Start the Arbitrary waveform editor.

Select

Non Built-in Waveform from the Arbitrary’s drop down

list

and click

to start the Arbitrary

waveform editor.

5 Select the interpolation method.

Click

to enable or disable linear interpolation between

the user

-defined points. With interpolation enabled

(default), the waveform editor makes a straight

-line

connection between the points. With interpolation
disabled, the waveform editor maintains a constant
voltage level between the points and creates a “step

-like’

waveform.