Measurement output, Starting the continuous reading output, Stopping the continuous reading output – Vaisala MMT310 User Manual

Page 38: Set continuous output interval for run mode

Advertising
background image

User's Guide _______________________________________________________________________

38 ___________________________________________________________________ M210474EN-C

Measurement Output

Starting the Continuous Reading Output

Syntax: R <cr>

Starts output of measurements to the peripheral devices (RUN mode).

While the measurement output is active the only command that can be

given is S (stop).

The output mode can be changed with the FORM command.

Example:

>r

aw= 0.2000 aw T= 25.09 'C …

aw= 0.2000 aw T= 25.20 'C …

Stopping the Continuous Reading

Output

Syntax: S <cr>

Stops the continuous output. <ESC> can also be used to stop the output.

Set Continuous Output Interval for RUN

Mode

Syntax: INTV xxx yyy<cr>

Where
Xxx = Output interval (0 ... 255)

yyy = Unit (s, min or h)

Example:

>intv 1

Output interval: 1 S

>intv 1 min<cr>

Output interval: 1 MIN

>intv 1 h<cr>

Output interval: 1 H