High end adjustment – Vaisala MMT310 User Manual

Page 58

Advertising
background image

User's Guide _______________________________________________________________________

58 ___________________________________________________________________ M210474EN-C

High End Adjustment

6.

After you have performed the low end adjustment, insert the probe

head into a measurement hole of the high end reference chamber

(e.g. NaCl: 75 % RH chamber in the humidity calibrator HMK15,

use the adapter fitting (13.5 mm)). Note that the difference between

the two humidity references must be at least 50% RH.

7.

Let the probe stabilize at least 30 minutes. You can follow the

stabilization by pressing any key, typing C and pressing ENTER.

8.

When stabilized, type the high end reference value after the

question mark and press ENTER.

crh

RH : 11.25 Ref1 ? c

RH : 11.24 Ref1 ? c

RH : 11.24 Ref1 ? 11.3

Press any key when ready ...

RH : 75.45 Ref2 ? c

RH : 75.57 Ref2 ? c

RH : 75.55 Ref2 ? c

RH : 75.59 Ref2 ? 75.5

OK

>
OK

indicates that the calibration has succeeded

9.

Take the probe out of the reference conditions and replace the

filter. Take care to tighten the filter properly, recommended force

130 Ncm.

10. If needed, give the calibration information (date and text) to the

transmitter's memory, see the serial commands on page 61.

11. Reset the transmitter by giving the RESET command. The

transmitter returns to normal mode.