Analog Output Calibration

Syntax: ACAL<cr>

Connect the MMT310 to a multimeter in order to measure either current

or voltage depending on the selected output type. Issue the ACAL

command. Then type the multimeter reading and press Enter.

Example (current outputs):

>acal

Ch1

I1 (mA) ?

2.046

Ch1

I2 (mA) ?

18.087

Ch2

I1 (mA) ?

2.036

Ch2

I2 (mA) ?

18.071

>

Adjustment Using an MI70 Indicator

You can use the MI70 handheld indicator to adjust the following in

MMT310:

- Temperature (T)
- Relative humidity (RH)
- Pressure compensation value for humidity measurement (environment

settings)

Note that starting the adjustment mode in MI70 requires pressing the

adjustment button on the MMT310 board. See Figure 15 on page 56 for

the location of the adjustment button.

NOTE

When adjusting both relative humidity and temperature, make the

temperature adjustment first.

Starting MI70 Adjustment

1.

Connect the transmitter to the MI70 hand-held indicator using the

connection cable (optional accessory, Vaisala order code

DRW216050SP).

2.

Turn on the MI70.

3.

Press the MMT310 adjustment button (see Figure 15 on page 56)

to start the adjustment mode. A notification about starting the

adjustment mode is displayed on the MI70 screen.

4.

Press OK to enter the adjustment menu. The available adjustment

options are displayed (RH or T). You can also view the date of last

adjustment.

Advertising