Chapter 5 _______________________________________________________________ Maintenance

VAISALA ________________________________________________________________________ 53

Clean the Sensor Before Storing the

Transmitter

Clean the used sensor with instrument air to blow out existing oil or

gently first flush with heptane (C7H16) and dry with instrument air to

prevent oxidation of the sensor. The oxidation of the sensor can cause

extended response times or drifting.

Technical Support

For technical questions, contact Vaisala technical support by e-mail at

[email protected]

. Provide at least the following supporting

information:

- Name and model of the product in question
- Serial number of the product
- Name and location of the installation site
- Name and contact information of a technically competent person who

can provide further information on the problem.

Product Returns

If the product must be returned for service, see

www.vaisala.com/returns

.

For contact information of Vaisala Service Centers, see

www.vaisala.com/servicecenters

.