Revert factory calibration, View current offset and gain adjustment – Vaisala MMT310 User Manual

Page 60

Advertising
background image

User's Guide _______________________________________________________________________

60 ___________________________________________________________________ M210474EN-C

Revert Factory Calibration

Syntax: LI<cr>

1.

Remove the transmitter unit from the mounting (see page 18) and

press the adjustment button once (see Figure 15 on page 56).

2.

Issue the LI command and set the offset value to 0 and gain value

to 1.

>li

RH offset : -0.6000000 ? 0

RH gain : 1.00000000 ? 1

T offset : 0.00000000 ? 0

T gain : 0.40000000 ? 1

>

3.

Reset the transmitter by giving the RESET command.

The transmitter returns to normal mode.

View Current Offset and Gain Adjustment

View the currently applied standard user adjustment offset and gain with

the L command.

Syntax: L<cr>

Example:

>l

RH offset : 0.00000000

RH gain : 1.00000000

T offset : 0.00000000

T gain : 1.00000000

>