Changing the sensor, Figure 14 – Vaisala MMT310 User Manual

Page 52

Advertising
background image

User's Guide _______________________________________________________________________

52 ___________________________________________________________________ M210474EN-C

When replacing the filter, wear clean gloves to avoid depositing dirt or

oil on the filter.

Replace the filter as follows:

1.

Turn the filter counterclockwise until it is loose.

2.

Pull the filter straight out carefully. Do not damage the sensor.

3.

Take the new filter, and insert it to the filter thread.

4.

Tighten the new filter to 5 Nm by turning it clockwise.

Changing the Sensor

NOTE

When replacing the sensor, the new sensor must be of the same type as

the old sensor (HUMICAP180L2 for MMT310). Information on Vaisala

sensors is available online at

www.vaisala.com/sensorinfo

.

Changing the sensor should be considered corrective maintenance, and it

is not necessary in normal operation. If the accuracy of the transmitter

does not seem to be within specification, it is more likely that the

transmitter is in need of calibration and adjustment, and not sensor

replacement. Refer to section Calibration and adjustment on page 55.

1.

Remove the filter from the probe. See the instructions in section

Changing the Filter on page 51.

2.

Remove the damaged sensor and insert a new one. Handle the new

sensor by the plastic socket. DO NOT TOUCH THE SENSOR

PLATE.

3.

Perform a calibration and adjustment as instructed in section

Relative Humidity Calibration after Sensor Change on page 61.

4.

Attach a new filter on the probe. When using the stainless steel

filter, take care to tighten the filter properly (recommended force

5 Nm).

0508-079

Figure 14

Changing the Sensor

The following numbers refer to Figure 14 above:
1 = Pull out the sensor

2 = Plastic socket