Chapter 6 ____________________________________________________ Calibration and adjustment

VAISALA ________________________________________________________________________ 61

Relative Humidity Calibration after

Sensor Change

Syntax: FCRH<cr>

The transmitter asks and measures relative humidity and calculates the

calibration coefficients. This two-point adjustment needs to be performed

after a sensor change. Follow the more detailed calibration instructions

on page 56, but instead of the CRH command, use the FCRH command.

>FCRH

RH

:

1.82 1. ref

?

0

Press any key when ready...

RH

: 74.22

2. ref

? 75

OK

>
OK

indicates that the calibration has succeeded.

Set Calibration Information Text

Syntax: CTEXT<cr>

1.

Remove the transmitter unit from the mounting (see page 18) and

press the adjustment button once (see Figure 15 on page 56).

2.

Issue the LI command and enter value 0 for offset and value 1 for

gain.

3.

Issue the CTEXT command and enter calibration information text.

Set Calibration Date

Syntax: CTEXT<cr>

1.

Remove the transmitter unit from the mounting plate (see section

Mounting the Transmitter/Removing the Transmitter Unit on page

18) and press the adjustment button once (see Figure 15 on page

56).

2.

Issue the LI command and enter value 0 for offset and value 1 for

gain.

3.

Issue the CDATE command and set the calibration date.

Example:

CDATE 2001 12 11

Calibration : 2001-12-11

>