Environment settings, Last adjustment date – Vaisala MMT310 User Manual

Page 66

Advertising
background image

User's Guide _______________________________________________________________________

66 ___________________________________________________________________ M210474EN-C

4.

Give the higher reference temperature by using the arrow buttons

and press

OK

.

5.

To confirm the adjustment, press

YES.

To cancel entering the

adjustment, press

NO

to return to adjustment mode display

without making any changes.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the T adjustment mode and return to the main adjustment options.

Press

EXIT

to quit the adjustment mode and return to the basic MI70

view.

Environment Settings

When you select RH adjustment for the first time from the adjustment

main menu, MI70 asks if you want to check the environment settings

(pressure compensation value for humidity measurement).

If you need to access the environment settings after this, exit the

adjustment mode by selecting EXIT in the main adjustment menu and

navigate to the environment menu from the MI70 main menu. To return

to the adjustment mode, press the adjustment button on the MMT310

board.

Last Adjustment Date

To see the last adjustment date, select Last adjustment date in the main

adjustment menu. You can also view this date from the Device

information menu in the MI70 main menu.