Select parameter for analog outputs, Table 13, Analog output quantity abbreviations – Vaisala MMT310 User Manual

Page 45

Advertising
background image

Chapter 4 _________________________________________________________________ Operation

VAISALA ________________________________________________________________________ 45

NOTE

1 … 5 V output can be selected when ordering the device. The 1 .. 5

output is based on the 0 … 5 V output mode, but has been adjusted to

1 … 5 V by scaling the valid range to 1 … 5. In the scaled 1 … 5 V

output mode, 0 V is reserved for error output (AERR set to 0).

If the output includes the reminder 'remember to set jumpers', ignore the

reminder.

Select Parameter for Analog Outputs

Syntax: ASEL xxx yyy<cr>

Where
xxx =

Quantity of channel 1

yyy =

Quantity of channel 2

After entering the command, the low and high values of each channel are

displayed one row at a time and a new value can be entered. Enter a new

value and press <cr> to move to the next row, or press <cr> without

entering a value to keep the current setting.

Use abbreviations shown in Table 13 Analog Output Quantity

Abbreviations below.

Table 13

Analog Output Quantity Abbreviations

Quantity

Abbreviation

aw water activity

-

T Temperature

°

C

ppm (for

transformer oil only)

ppm

Examples:

>asel aw t

Ch1 aw lo : 0.00 ?

Ch1 aw hi : 1.00 ?

Ch2 T lo : -40.00 'C ?

Ch2 T hi : 60.00 'C ?

>asel t ppm

Ch1 T lo : -40.00 'C ?

Ch1 T hi : 60.00 'C ?

Ch2 ppm lo : 0.00 ppm ?

Ch2 ppm hi : 5000.00 ppm ?

>