Other commands, Check transmitter settings – Vaisala MMT310 User Manual

Page 48

Advertising
background image

User's Guide _______________________________________________________________________

48 ___________________________________________________________________ M210474EN-C

Other Commands

Check Transmitter Settings

Syntax: ?<cr>

Syntax: ??<cr>

Use the ? command to check the current transmitter configuration. The

?? command is similar but can also be used if the transmitter is in POLL

mode.

Example (factory default settings):

>?

MMT310 / 1.03

PRB serial nr : V1234567

Calibration : 2003-03-25

Cal. info : NU/HMK15

Output units : metric

Pressure : 1013.25 hPa

RS232 settings

Address : 2

Output interval: 1 MIN

Baud P D S : 4800 E 7 1

Serial mode : STOP

Analog outputs

Ch1 output mode: 0 ...20mA

Ch2 output mode: 4 ...20mA

Ch1 error out : 0.000mA

Ch2 error out : 0.000mA

Ch1 T lo : -40.00 'C

Ch1 T hi : 60.00 'C

Ch2 ppm lo : 0.00 ppm

Ch2 ppm hi : 5000.00 ppm

>