Appendix A ______________________________________________________________ Applications

VAISALA ________________________________________________________________________ 79

Technical Data

Typical measuring range

0...80* ppm (0...100ºC)

*Upper edge limited to saturation

Accuracy

(Calculation Model with

Average Coefficients)

Temperatures > 30 ºC: better than 10 % of

the reading

Temperatures < 30ºC: see Figure 23 below.

1311-206

Figure 23

Maximum Errors Caused by Deviation of Mineral Oils

Using Calculation Model with Average Coefficients

Temperature

Measurement range

-40...+180ºC

Response times (with stainless steel filter)

In still air (20ºC)

10 s

In still transformer oil

(20ºC)

< 10 min