Output formatting, Set serial output format, Table 11 – Vaisala MMT310 User Manual

Page 42: Modifiers, Table 12, Analog output quantity abbreviations

Advertising
background image

User's Guide _______________________________________________________________________

42 ___________________________________________________________________ M210474EN-C

Output Formatting

Set Serial Output Format

Syntax: FORM x<cr>

Where
x = Formatter string

The FORM command can be used to change the output format of the

SEND and R commands.

A formatter string consists of quantities and modifiers: use the modifiers

presented in Table 11 and quantities presented in Table 12 when

configuring the output format.

Use the FORM / command to return the default output form.

Table 11

Modifiers

Modifier

Description

x.y

Length modifier (whole numbers and decimal places)

#t

Tabulator

#r

Carriage return

#n

Line-feed

""

String-constant

U5

Unit field and length (unit can be changed with the UNIT

command)

Table 12

Analog Output Quantity Abbreviations

Quantity

Abbreviation

aw water activity

aw

T Temperature

T

ppm (for transformer oil only)

ppm

Example:

>form "aw="1.2 aw #r#n

aw=0.79

>form "aw="1.2 aw U3 #t "ppm="3.1 ppm " " U5 #t "T="3.1 T

U3 #r#n

aw=0.78 ppm= 53.4 ppm T= 27.3'C

>

>send

2000-01-01 20:56:27 N 0 aw=0.78 ppm= 53.2 ppm T=

27.2'C

>