Licl – nacl automatic adjustment – Vaisala MMT310 User Manual

Page 64

Advertising
background image

User's Guide _______________________________________________________________________

64 ___________________________________________________________________ M210474EN-C

5.

To confirm the adjustment, press

YES.

To cancel entering the

adjustment, press

NO

to return to the adjustment mode display

without making any changes.

Note that if the difference between the two references is less than

50 %, adjustment cannot be done.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the RH adjustment mode and return to the main adjustment options.

Press

EXIT

to quit the adjustment mode and return to the basic MI70

view.

LiCl – NaCl Automatic Adjustment

LiCl-NaCl automatic adjustment is a two-point adjustment in reference

humidities of 11.3 % (LiCl) and 75 % (NaCl). You do not have to enter

the reference values, the MI70 displays the accurate value based on the

measured temperature and the Greenspan table stored into the memory of

the MI70.

To make the adjustment, select LiCl – NaCl auto in the RH adjustment

menu and follow the steps below:

1.

Set the probe to the LiCl reference humidity. You can follow the

stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized.

2.

Set the probe to the NaCl reference humidity. You can follow the

stabilization from the

GRAPH

display. Press

READY

when the

reading is stabilized.

3.

After the NaCl reference humidity measurement is done, MI70

checks if you want to use this 2-point adjustment. Press

OK

to

confirm the adjustment.

Once confirmed, the adjustment has been carried out. Press

BACK

to

exit the RH adjustment mode and return to the main adjustment options.

Press

EXIT

to quit the adjustment mode and return to the basic MI70

view.