Scale analog outputs, Test analog outputs – Vaisala MMT310 User Manual

Page 46

Advertising
background image

User's Guide _______________________________________________________________________

46 ___________________________________________________________________ M210474EN-C

Scale Analog Outputs

Syntax: ASCL<cr>

Example:

>ascl

Ch1 T lo : -40.00 'C ? -20

Ch1 T hi : 60.00 'C ? 40

Ch2 ppm lo : 0.00 ppm ? 0

Ch2 ppm hi : 5000.00 ppm ? 3000

>ascl

Ch1 T lo : -20.00 'C ?

Ch1 T hi : 40.00 'C ?

Ch2 ppm lo : 0.00 ppm ?

Ch2 ppm hi : 3000.00 ppm ?

The ppm calculation is only used for transformer oils.

Test Analog Outputs

Syntax: ITEST aa.aaa bb.bbb<cr>

The operation of the analog outputs is tested by forcing the outputs to

given values. The values in the analog outputs can then be measured with

a current/voltage meter.

Where
aa.aaa

= Current/voltage value to be set for channel 1

bb.bbb

= Current/voltage value to be set for channel 2

Example:

>itest 8.30 6.40

8.30000mA 2c30 6.40000mA 2278

>

The ITEST command output includes internal debugging values shown

in hexadecimal format (

2c30

and

2278

in the example above).

The set values remain valid until you issue the ITEST command without

readings or reset the transmitter.

Advertising