Reset transmitter, Display error messages – Vaisala MMT310 User Manual

Page 50

Advertising
background image

User's Guide _______________________________________________________________________

50 ___________________________________________________________________ M210474EN-C

Reset Transmitter

Syntax: RESET<cr>

This command resets the device, resulting in a short start-up delay. The

serial interface mode (STOP, RUN or POLL) selected with the SMODE

command is taken into use at restart.

Display Error Messages

Syntax: ERRS<cr>

Display transmitter error messages. If there are no errors present,

a PASS will be returned.

Examples:

>errs

PASS

>

>errs

FAIL

Error: Temperature measurement malfunction

Error: Humidity sensor open circuit

>

In case of a constant error, please contact a Vaisala Service Center.

For information on contacting Vaisala Service Centers, see

www.vaisala.com/servicecenters

.