Preparations Before the Calibration

Before calibration, clean the used sensor with instrument air to blow out

existing oil, or gently first flush with heptane (C7H16) and dry with

instrument air to decrease response time.

NOTE

It is important to clean the sensor before calibration as a dirty sensor can

contaminate the salt bath and change the reference condition.

Relative Humidity Calibration and Adjustment (in

Two Points)

Relative humidity calibration and adjustment of the MMT318 and

MMT317 is done by using two reference humidities (low end and high

end). The difference between the two humidity references must be at

least 50% RH.

An easy calibration can be done by using Vaisala Humidity Calibrator

HMK15. When using the HMK15 calibrator, use the adapter fitting (13.5

mm) with the MMT318 and MMT317 probe.

Before calibration the transmitter must be set to adjustment mode by

pressing the adjustment button (see Figure 15 below).