Testing solenoids and relays, Fig. 15 – Actron Digital Multimeter CP7676 User Manual

Page 14

Advertising
background image

View reading on display:
• Switch is closed if display reading is less

than 10

.

• Switch is open if display reading

indicates an overrange condition. (see

Setting the Range on page 6)

7. Operate switch.

View reading on display:

• Switch is closed if display reading is less

than 10

.

• Switch is open if display reading indicates

an overrange condition. (see Setting the

Range on page 6)

8. Repeat Step 7 to verify switch

operation.

Good Switch: Display reading alternates

from a 10

or less value to an overrange

condition as you operate switch.

Bad Switch: Display reading remains

unchanged as you operate switch.

Testing Solenoids and Relays

This test checks to see if a solenoid or relay

have a broken coil. If the coil tests good, it is

still possible that the relay or solenoid are

defective. The relay can have contacts that

are welded or worn down, and the solenoid

may stick when the coil is energized. This test

does not check for those potential problems.

To test Solenoids and Relays (see Fig. 15):

14

1. Insert BLACK test lead into the COM

test lead jack.

2. Insert RED test lead into the

V

test

lead jack.

3. Turn multimeter rotary switch to 200

range.

Most solenoids and relay coil resistances

are less than 200

. If meter overranges,

turn multimeter rotary switch to next

higher range. (see Setting the Range on

page 6)

4. Touch RED and BLACK test leads

together and view display.

Display should read typically 0.2

to 1.5

.

If display reading was greater than 1.5

,

check both ends of test leads for bad

connections. If bad connections are

found, replace test leads.

5. Connect BLACK test lead to one side

of coil.

6. Connect RED test lead to other side of

coil.

7. View reading on display.

• Typical solenoid / relay coil resistances

are 200

or less.

• Refer to vehicle service manual for your

vehicles resistance range.

8. Test Results

Good Solenoid / Relay Coil: Display in

Step 7 is within manufacturers

specification.

Bad Solenoid / Relay Coil:

• Display in Step 7 is not within

manufacturers specifications.

• Display reads overrange on every ohms

range indicating an open circuit.

NOTE: Some relays and solenoids have

a diode placed across the coil. To test this

diode see Testing Diodes on page 9.

2K 200

20M

2M

200K

20K

OHMS

750V AC
200V DC

COM

V

DWELL

4

CYL

5

CYL

6

CYL

8

CYL

Fig. 15

Relay or Solenoid

Red

Black