Myron L PS9TK User Manual

Page 15

Advertising
background image

11

the display and holding either key scrolls the value rapidly.
While in calibration or calculator mode, the keys step or scroll the

displayed value up or down.
While in Memory Recall, the keys scroll the display up and down

through the stack of records (ref. Memory Recall, pg. 37).

iv.

aFTER uSing ThE P

OOL

P

RO

A. Maintenance of the Conductivity Cell

Rinse out the cell cup with clean water. Do not scrub the cell. For

oily films, squirt in a foaming non-abrasive cleaner and rinse (ref.

Cleaning Sensors, pg. 48). Even if a very active chemical discolors

the electrodes, this does not affect the accuracy; leave it alone.

B. Maintenance of the pH/ORP Sensor

The sensor well must be kept wet with a saline solution. Before

replacing the rubber cap, rinse and fill the sensor well with Myron L

pH Sensor Storage Solution. If unavailable, you can use an almost

saturated KCl solution, pH 4 buffer or a saturated solution of table salt

and tap water (ref. pH and ORP/Free Chlorine Practices to Maintain

Calibration, pg. 37). nEvER uSE DiSTiLLED waTER.
if all ORP solutions are reporting approximately the same value,

clean OnLy the platinum ORP electrode with an MLc ORP

conditioner-soaked cotton swab, being careful not to touch the

swab to the glass bulb of the ph sensor.

v.

SPEciFic REcOMMEnDED MEaSuRing PROcEDuRES

A. Parameter Methods

For Conductivity, Mineral/Salt and Total Dissolved Solids measurements,

if the proper solution is not selected (KCl, NaCl or 442), see Solution

Selection, pg. 24.

nOTE: After sampling high concentration solutions or temperature

extremes, more rinsing may be required. When sampling low

conductivity solutions, be sure the pH cap is well seated so that no

solution washes into the conductivity cell from around the pH cap.

B. Titration Methods

For Alkalinity, Hardness titrations, assemble the required materials

before you begin: a pipette with 100µL tips, the required reagents, the

cell extender, and the T-plunger.

Advertising