Xii. TOTaL RETuRn to FacTORy SETTingS “Fac SEL”

There may come a time when it would be desirable to quickly reset all

the recorded calibration values in the instrument back to the factory

settings. This might be to ensure all calibrations are set to a known

value, or to give the instrument to someone else free of adjustments

or recorded data for a particular application.
nOTE: all stored data will be lost.
1.

Press

.

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the “cL

r

aLL” and the

C-F locations. The display will show a “Fac SEL” (see Figure 17).

4.

Press

to accept the resetting. Display will return to

Conductivity.