Myron L PS9TK User Manual

Page 61

Advertising
background image

57

Clearly, the solution characteristics should be chosen to truly represent

the actual water under test for rated accuracy of ±1%. Many industrial

applications have historically used relative measurements seeking

a number to indicate a certain setpoint or minimum concentration

or trend. The P

OOL

P

RO

gives the user the capability to collect data

in “KCl conductivity units” to compare to older published data, in

terms of NaCl or 442, or as appropriate. The P

OOL

P

RO

can be used

to reconcile data taken with other compensation assumptions.

XX.

cOnDucTiviTy cOnvERSiOn to TOTaL DiSSOLvED

SOLiDS (TDS)

Electrical conductivity indicates solution concentration and ionization

of the dissolved material. Since temperature greatly affects ionization,

conductivity measurements are temperature dependent and are

normally corrected to read what they would be at 25°C (ref. Temperature

Compensation, pg. 58).

A. How it’s Done

Once the effect of temperature is removed, the compensated conductivity

is a function of the concentration (TDS). Temperature compensation of

the conductivity of a solution is performed automatically by the internal

processor with data derived from chemical tables. Any dissolved salt at

a known temperature has a known ratio of conductivity to concentration.

Tables of conversion ratios referenced to 25°C have been published by

chemists for decades.

B. Solution Characteristics

Real world applications have to measure a wide range of materials and

mixtures of electrolyte solutions. To address this problem, industrial users

commonly use the characteristics of a standard material as a model for

their solution, such as KCl, which is favored by chemists for its stability.

Users dealing with seawater, etc., use NaCl as the model for their

concentration calculations. Users dealing with freshwater work with

mixtures including sulfates, carbonates and chlorides, the three

predominant components (anions) in freshwater that the Myron L

Company calls “natural water”. These are modeled in a mixture called

“442™” which the Myron L Company markets for use as a calibration

standard, as it does standard KCl and NaCl solutions.

The P

OOL

P

RO

contains algorithms for these 3 most commonly referenced

compounds. The solution type in use is displayed on the left: KCl, NaCl,

and 442.