Myron L PS9TK User Manual

Page 25

Advertising
background image

21

of table salt and tap water (ref. Cleaning Sensors, 2. pH/ORP,

pg. 48). Do not allow pH/ORP sensor to dry out.
H. Measuring ORP

The P

OOL

P

RO

features the ability to measure the activity of oxidizing or

reducing chemicals in solution as ORP mV. The instrument also includes

an innovative Free Chlorine Equivalent (FC

E

) feature (Measuring Free

Chlorine Using FC

E

, pg. 22) that uses ORP and pH to measure free

available chlorine (FAC) concentration in ppm. ORP mV and ppm of

free available chlorine (FAC) are the two most commonly used sanitizer

units of measure in water quality management.

1. ORP / FC

E

Mode Selection

The P

OOL

P

RO

allows the user to choose between measuring oxidizing

sanitizers using either ORP mV or as parts per million (ppm) of equivalent

free chlorine. Use ORP to directly measure the oxidizing power of all

sanitizers like ozone, bromine, peracetic acid or chlorine. Use FC

E

to

measure the strength of oxidizing sanitizers as ppm of equivalent free

chlorine. To select between ORP and Free Chlorine modes:
1.

Press

.

2.

Press and hold

for approximately 3 seconds.

The current preference for ORP units of measure is displayed. Factory

setting for this preference is ORP mV.

3.

Press the

or

keys to toggle between mV (standard

ORP mode) and FC

E

ppm. The setting chosen is displayed.

PPM

4.

Press any parameter key to exit ORP unit preference selection

or let the PS9 time out. ORP unit preference will be saved.