Xiii.

cELL chEcK

The cell check verifies the cleanliness of the conductivity/MIN/

SALT/TDS sensor. In normal use the cell may become dirty or

coated and require cleaning. If the display is showing “.00” when

the cell cup is dry, the sensor is probably clean. No matter what a

manufacturer claims, a sensor can and will become contaminated or

coated and, therefore, require cleaning. A true 4-wire sensor, as in

the P

OOL

P

RO

, helps to mitigate contamination, but NO SENSOR IS

100% IMMUNE.

1.

Press

.

2.

Press

to display the stored memory records.

3.

Press

repeatedly until you pass the FAC SEL location.

The display will show a “cELL ch” (see Figure 18).

4.

Press

to test. If cell is clean, “good” will momentarily be

displayed (see Figure 19). If cell is dirty, “cELL cLn” will be

displayed (see Figure 20) (ref. Cleaning Sensors, pg. 48).

Xiv. auTO OFF

Auto off allows the user to adjust the time the instrument is ON (up to

75 seconds) after each press of a key. Default time is 15 seconds with

60 seconds in “caL” (calibration) mode.

1.

Press

.

Figure 18

Figure 19

Figure 20