and holding the PS9 as prompted, then pressing caL as

prompted.

A hardness value will display when the titration measurement

is complete.

24.

Press caL. The “caL” icon will display.

25.

Adjust the displayed value to “200” using UP or DOWN key. The

PS9 will not allow you to scroll up or down to “200” if it is beyond

±10% of the reading. Ref. Total Return to Factory Settings “Fac

SEL”, pg. 42 to reset the Hardness Calibration to the “Fac

setting along with all other calibrations.

26.

Press caL to accept.

The PS9 will automatically power off after the period of inactivity

defined in the AUTO OFF setting (ref. AUTO OFF, pg. 43).

5. pH Calibration

iMPORTanT: Always “zero” your P

OOL

P

RO

with a pH 7 buffer solution

before adjusting the gain with acid or base buffers, i.e., 4 and/or 10,

etc.

a. pH Zero Calibration

1.

Rinse sensor well and cell cup 3 times with 7 buffer solution.

2.

Refill both sensor well and cell cup with 7 buffer solution.

3.

Press

to verify the pH calibration. If the display shows

7.00, skip the pH Zero Calibration and proceed to section b.

pH gain Calibration.

4.

Press

to enter calibration mode. The “caL”, “buFFER

and “7” annunciators will appear (see Figure 3). Displayed

value will be the uncalibrated sensor.

nOTE: If a wrong buffer is added (outside of 6-8 pH), “7”and “buFFER

will flash, and the P

OOL

P

RO

will not adjust. The uncalibrated pH value

displayed in step 4 will assist in determining the accuracy of the pH