Myron L PS9TK User Manual

Page 39

Advertising
background image

35

nOTE: If the “acd” and “baS” indicators are blinking, the PS9 is indicating

an error and needs either an acid or base solution present in the sensor

well.

3.

Rinse sensor well 3 times with acid or base buffer solution.

4.

Refill sensor well again with same buffer solution.

5.

Press

or

until display agrees with buffer value.

6.

Press

to accept 2nd point of calibration. Now the

display indicates the next type of buffer to be used.

Single point gain Calibration is complete. You may continue for the 3rd

point of Calibration (2nd gain) or exit by pressing any measurement key.

Exiting causes the value accepted for the buffer to be used for both acid

and base measurements.

To continue with 3rd point calibration, use basic buffer if acidic buffer

was used in the 2nd point, or vice-versa. Again, match the display to

the known buffer value as in step 2 and continue with the following

steps:

7.

Repeat steps 3 through 6 using opposite buffer solution.

8.

Press

to accept 3rd point of calibration, which

completes the Calibration procedure. Fill sensor well with

Myron L Storage Solution and replace protective cap.

6. ORP Calibration

ORP electrodes rarely give false readings without problems in the

reference electrode. For this reason, and because calibration solutions

for ORP are highly reactive and potentially hazardous, your P

OOL

P

RO

has an electronic ORP calibration. This causes the zero point on the

reference electrode to be set whenever pH 7 calibration is done.

7. Temperature Calibration

Temperature calibration is not necessary in the P

OOL

P

RO

.

Advertising