Myron L PS9TK User Manual

Page 33

Advertising
background image

29

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

Alkalinity Standard 100PPM.

4.

Refill cell cup and cell extender with Alkalinity Standard 100PPM.

5.

Insert T-plunger into open cell extender until the “T” rests flush

on the lip of the cell extender. This will cause fluid beyond what

is required for titration to overflow the cell extender.

6.

Rotate the T-plunger from side to side, allowing the arms of the

T-plunger to slide along the rim of the cell extender.

7.

Carefully remove the T-plunger so as not to flick or spill remaining

sample. Tap the T-plunger on the side of the cell extender to

remove sample that may be stuck to the T-plunger.

8.

Press

. “aL alternating with “tdS” briefly displays along

with the “PPM” value of the sample. “P

r

ES caL” will display

when the reading stabilizes.

9.

Press caL to accept and advance to the first titration prompt.

add” alternating with “a1” displays; then “agit t1” alternating

with “hOLd” displays. (The number after “t” indicates which

titration point you are on. The number “1” here indicates this is

the first titration point. Subsequent titration points are sequenced

numerically: 2, 3, 4, etc.)