Myron L PS9TK User Manual

Page 31

Advertising
background image

27

To bypass a calibration, simply press

to accept the present

value as-is.

b. Leaving Calibration

Pressing

accepts calibration and exits CAL mode if there are no

more steps. Calibration is complete when the “caL” icon goes out.

Pressing any measurement key during calibration cancels changes

not yet accepted and exits calibration mode. Leaving pH after the 2nd

buffer results in the same gain being entered in place of the 3rd buffer.

2. Calibration Limits

There are calibration limits for all functions for which calibration is

required. For Alkalinity and Hardness, the PS9 will not allow you to

calibrate beyond its calibration limits. For all other functions, attempts

to calibrate too far up or down from the ideal “Fac” value stored in the

PS9 by the factory will cause the displayed value to be replaced with

Fac”. If you accept it (press the caL key), you will have the original

default factory calibration for this measurement. The need to calibrate

so far out that “Fac” appears indicates a procedural problem, incorrect

standard solution, a very dirty cell cup or an aging pH/ORP sensor (ref.

Troubleshooting Chart, pg. 50).

C. Calibration Procedures

1. Conductivity, MIN/SALT or TDS Calibration

a.

Rinse conductivity cell 3 times with proper standard (KCl,

NaCl, or 442) (ref. Cond/TDS Standard Solutions, pg. 52).

b.

Refill conductivity cell with same standard. KCl-7000 shown.

c.

Press

,

MIN

SALT

or

, then press

; “caL” icon

will appear on the display (see Figure 2).

d.

Press

or

to step the displayed value toward the

Figure 2