For maximum accuracy, keep the pH sensor cap on tight so

that no fluid washes into the conductivity cell.

D. pH and ORP/Free Chlorine Practices to Maintain Calibration

1.

Keep the sensor wet with Myron L Storage Solution.

2.

Rinse away caustic solutions immediately after use.

ORP calibration solutions are caustic, and ±5% is considered very

accurate. By using the pH zero setting (0 mV = 7 pH) for ORP and

precision electronics for detection, the P

OOL

P

RO

delivers better

accuracy without calibration than a simpler instrument could using

calibration solutions.

iX.

MEMORy

This feature allows up to 100 readings with their temperatures to be

stored simultaneously for later recall. At the same time, the TIME and

DATE are also recorded. To download the memory to a computer, ref.

bluDock™ WIRELESS DATA TRANSFER INSTRUCTIONS, pg. 45.

A. Memory Storage

1.

While displaying a measurement, press

to record the

displayed value.

2.

MEMORy” will appear and the temperature display will

be momentarily replaced by a number (1-100) showing the

position of the record. Figure 6 shows a reading of 1806 µS

stored in memory record #4.

B. Memory Recall

1.

Press any measurement key.

2.

Press

; “MEMORy” will appear, and the display will show

the last record stored.

3.

Press

or

to scroll to the record location desired

Figure 6