Myron L PS9TK User Manual

Page 42

Advertising
background image

38

(the temperature display alternates between temperature

recorded and location number).

4.

Press

to display time and date stamp.

5.

Press any measurement key to leave memory recall or allow to

automatically turn off.

C. Clearing a Record/Memory Clear

After recalling a certain record location, press and HOLD

to

clear that memory. This space will be the place for the next memory

record, unless you scroll to another empty position before ending the

recall sequence. The next memory stored will go into the next highest

available memory location.

Example: You have locations 1-7 filled and wish to clear the

conductivity reading stored in record location #3 and replace it with a

pH reading.

1.

Press

and scroll to location #3.

2.

Press and HOLD

to clear old record #3.

3.

Fill pH/ORP sensor well with sample.

4.

Press

to measure sample and press

to store

reading in location #3.

5.

The next memory stored will go into location #8.

6.

To clear all records: After pressing

, scroll down. “cL

r

aLL” will be displayed (see Figure 7).

7.

Press

. All records will be cleared.

Figure 7