Reference Junction, is of the same construction with a porous plug

in place of a glass barrier to allow the junction fluid to contact the

test solution without significant migration of liquids through the plug

material. Figure 28 shows a typical 2 component pair. Migration does

occur, and this limits the lifetime of a pH junction from depletion of

solution inside the reference junction or from contamination. The

junction may be damaged if dried out because insoluble crystals may

form in a layer, obstructing contact with test solutions.

Glass surface

Figure 30

KCl solution

Electrode wire

Electrode

wire

H

+

ions

Junction

Plug

4. The Myron L Integral pH Sensor

The sensor in the P

OOL

P

RO

(see Figure 29) is a single construction

in an easily replaceable package. The sensor body holds an oversize

solution supply for long life. The reference junction “wick” is porous to

provide a very stable, low permeable interface, and is located under

the glass pH sensing electrode. This construction combines all the

best features of any pH sensor known.

KCl solution

Figure 31

Junction plug

Platinum button

H

+

ions

Electrode wires

Glass

Glass

Surface

Figure 28

Figure 29