Myron L PS9TK User Manual

Page 34

Advertising
background image

30

10.

Using the pipette, add 100 µL of Reagent: A1, close the cell

extender cap securely, then agitate and hold. Repeat as

prompted. “P

r

ES caL” will display.

11.

Press caL to advance to the next titration point. “add

alternating with “a1” displays; then “agit t2” alternating with

hOLd” displays.

12.

Tap the cell extender cap to dispel any solution clinging to the

cap back into the cell. Carefully open the cell extender cap so

as not to spill any solution.

13.

Continue adding 100µL of Reagent A1, agitating and holding the

PS9 as prompted, then pressing caL as prompted. An alkalinity

value will display when the titration measurement is complete.

14.

Press caL. The “caL” icon will display.

15.

Adjust the displayed value to “100” using the UP or DOWN key.

The PS9 will not allow you to scroll up or down to “100” if it

is beyond ±10% of the reading. Ref. Total Return to Factory

Settings “FAC SEL”, pg. 42 to reset the Alkalinity Calibration to

the “FAC” setting along with all other calibrations.

16.

Press caL to accept. The PS9 will automatically power off after

the period of inactivity defined in the AUTO OFF setting (ref.

AUTO OFF, pg. 43).

4. Hardness Calibration

Prepare the materials required for titration: Hardness Standard Solution

200PPM, Reagent: C1, Reagent: H1, Reagent: H2, Reagent: H3, cell

extender, T-plunger, 100 µL pipette with a clean tip installed, and two

extra clean tips.

nOTE: If you do not complete each titration point within the 3-minute

timeout period, the PS9 exits the Hardness function. Press haRD to

start the titration process over from step 1. Between titrations, the PS9

will automatically power off after the period of inactivity defined in the

AUTO OFF setting (ref. AUTO OFF, pg. 43).

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

Hardness Standard 200PPM.