14

b. Hold

When the display says to hold (“hOLd”):

Hold the instrument steady in the horizontal

position while the instrument records a

measurement. Agitate and hold as many times

as prompted. Allow the display to cycle through

all screens, e.g., “EdTa”, “agit” and “hOLd”, at

least once. Continue holding steady until “P

r

ES

caL” displays.
Hold the PS9 steady before pressing caL to accept readings.

C. Measuring Conductivity Mineral/Salt Concentration &

Total Dissolved Solids (TDS)

1.

Rinse cell cup 3 times with sample to be measured. (This

conditions the temperature compensation network and prepares

the cell.)

2.

Refill cell cup with sample.

3.

Press

,

MIN

SALT

or

.

4.

Take reading. A display of [- - - -] indicates an overrange

condition.

D. Measuring Alkalinity

Prepare the materials required for titration: Reagent: A1, cell extender,

T-plunger, and 100 µL pipette with a clean tip installed.

nOTE: If you do not complete each titration point within the 3-minute

timeout period, the PS9 exits the Alkalinity function. Between titrations,

the PS9 will automatically power off after the period of inactivity defined

in the AUTO OFF setting (ref. AUTO OFF, pg. 43). Press aLK to start the

titration process over from step 1.

If you make a gross error in your titration method, e.g., you forget to add

sample to the cell, “Err” will display. Start the titration over from Step 1

to obtain a reading.

1.

Ensure pH protective cap is secure to avoid contamination.

2.

Install the cell extender by pushing the base of the cell extender

into the conductivity cell until it is fully seated.

3.

Rinse the conductivity cell cup and cell extender 3 times with

solution to be measured.

hOLd”: When “hOLd

displays, hold the PS9

still with bottom of case

parallel to the ground.