Myron L PS9TK User Manual

Page 29

Advertising
background image

25

C. Calibration of Each Solution Type

There is a separate calibration for each of the 4 solution types. Note

that calibration of a 442 solution does not affect the calibration of a

NaCl solution. For example: Calibration (ref. Conductivity, MIN/SALT

or TDS Calibration, pg. 27) is performed separately for each type

of solution one wishes to measure (ref. Conductivity/TDS Standard

Solutions, pg. 52).

D. Procedure to Select a Solution

nOTE: Check display to see if solution displayed (KCl, NaCl or 442)

is already the type desired. If not:

1.

Press

,

MIN

SALT

or

to select the parameter on

which you wish to change the solution type.

2.

Press and hold

key for 3 seconds to make “SEL

appear (see Figure 1). For demonstration purposes, all 3

solution types are shown simultaneously.

Figure 1

KCl

442

NaCl

3.

Use the

or

key to select type of solution desired

(ref. Solution Characteristics, pg. 57). The selected solution

type will be displayed: KCl, NaCl or 442.

4.

Press

to accept new solution type.