Anritsu S251B User Manual

Page 22

Advertising
background image

ON

OFF

Turns the Site Master on or off. When turned on, the system state at the last
turn-off is restored. If the

ESCAPE/CLEAR

key is held down, the factory preset

state is restored.

PRINT

Prints the current display to the selected printer via the RS232 serial port.

RECALL
DISPLAY

Recalls a previously saved trace from memory. When the key is pressed, a Re-
call Trace selection box appears on the display. Select a trace using the
Up/Down arrow key and press the

ENTER

key to implement.

To erase a saved trace, highlight the trace and select the

DELETE TRACE

softkey. To erase all saved traces, select the

DELETE ALL TRACES

softkey.

RECALL
SETUP

Recalls a previously saved setup from memory location 1 through 10. When the
key is pressed, a

RECALL SETUP

selection box appears on the display. Select a

setup using the Up/Down arrow key and press the

ENTER

key to implement.

Setup 0 recalls the factory preset state.

RUN

HOLD

When in the Hold mode, this key starts the Site Master sweeping and provides a
Single Sweep Mode trigger; when in the Run mode, it pauses the sweep. When
in the Hold mode, the hold symbol (page 3-23) appears on the LCD. Hold mode
conserves battery power.

SAVE
DISPLAY

Saves up to 200 displayed traces to non-volatile memory. When the key is
pressed,

TRACE NAME:

appears in the lower left of the display. Save the dis-

play with up to 16 alphanumeric characters for that trace name and press the

ENTER

key to implement.

SAVE
SETUP

Saves the current system setup to 1 of 10 internal non-volatile memory loca-
tions. When the key is pressed, a

SAVE SETUP

selection box appears on the dis-

play. Use the Up/Down arrow key to select a setup and press the

ENTER

key to

implement.

START
CAL

Starts the calibration in SWR, Return Loss, Cable Loss, insertion loss, insertion
gain, or DTF measurement modes. Follow the text in the lower left message area
that instructs you to do the following:

q

CONNECT OPEN TO RF out PORT

, Press

ENTER

The Site Master then measures the calibration “open” that you must attach to
the end of the test port or to the end of the test port cable attached to the RF
Out port.

q

CONNECT SHORT TO RF out PORT

, Press

ENTER

The Site Master then measures the calibration “short” that you must attach to
the end of the test port or to the end of the test port cable attached to the RF
Out port.

q

CONNECT LOAD TO RF out PORT

, Press

ENTER

The Site Master then measures the 50

W load that you must attach to the end

of the test port or to the end of the test port cable attached to the RF Out port.

3-4

Chapter 3 Functions and Operations