Chapter 4 measurements, Enter – Anritsu S251B User Manual

Page 69

Advertising
background image

Step 11. Connect the Open, Short, Loads, and through adapter to the test port extension

cable and press the

ENTER

key as prompted. Refer to the Calibration Procedure

earlier in this chapter for detailed calibration instructions.

NOTE:
The number of data points displayed can be changed by selecting

RES

in the

DTF AID table.

When the calibration is complete the results should be similar to those shown in Figure
4-18.

4-23

Chapter 4 Measurements