Range er rors 3-25, Range errors -25 – Anritsu S251B User Manual

Page 43

Advertising
background image

Range Errors

A listing of Range Error messages is given in Table 3-3.

3-25

Chapter 3 Functions and Operations

Error Message

Description

RANGE
ERROR:F1 > F2

The start (F1) frequency is greater than the stop (F2) frequency.

RANGE
ERROR:D1 > D2

The start (D1) distance is greater than the stop (D2) distance.

RANGE
ERROR:D2 >
DMax=xx.x m (ft)

The stop distance (D2) exceeds the maximum unaliased range. This
range is determined by the frequency span, number of points,
and relative propagation velocity:

MaximumUnaliased Range

dp V f

F

F

=

´

-

( .

) (

) (

)

1 5 108

2

1

(meters)

Where:

dp is the number of data points (128, 256, or 512)
V

f

is the relative propagation velocity

F2 is the stop frequency in Hz
F1 is the start frequency in Hz
Maximum Unaliased Range is in meters

Refer to Figure 3-16, page 3-27, for a plot of maximum range versus
frequency span.

RANGE ERROR:
TOP<=BOTTOM

The SWR scale parameter top value is less than or equal to its bottom
value.

RANGE ERROR:
TOP>=BOTTOM

The RL scale parameter top value is greater than or equal to its bottom
value.

CAL
INCOMPLETE

A complete open, short, and load calibration or a complete open,
short, load isolation and thru calibration must be performed before
calibration can be turned on.

DIST REQUIRES
F1 < F2

Valid distance to fault plots require a non-zero frequency span.

DIST REQUIRES
CAL

Distance-to-fault measurements do not provide usable data with CAL
OFF.

NO STORED
SWEEP AT THIS
LOCATION

Attempting to recall a display from a location that has not been previ-
ously written to. That is, the location does not contain stored sweep.

USE OPTIONS
MENU TO
SELECT A
PRINTER

Attempting to print a display with no printer selected. Select a printer,
then retry.

DISTANCE AND
CABLE
INSERTION LOSS
MODE ARE
INCOMPATIBLE

DTF measurements only display RL or SWR versus distance.

CANNOT ZERO
NO DETECTOR
INSTALLED

Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.

Table 3-3.

Range Error Messages (1 of 2)