Set ting the am pli tude scale 4-4, Saving the setup 4-4, Re calling a setup 4-4 – Anritsu S251B User Manual

Page 50: Setting the amplitude scale -4, Saving the setup -4, Recalling a setup -4

Advertising
background image

Step 15. Remove the Short, connect the calibrated Load and press the

ENTER

key. The

message “MEASURING LOAD ” will appear.

NOTE:
For OSLT calibration, the messages “MEASURING LOAD” and “CONNECT
LOADS TO RF OUT AND RF IN PORTS” will appear.

Step 16. Connect the second calibrated load to the RF In test port and press the

ENTER

key. The messages “MEASURING ISOLATION” and “CONNECT RF OUT
PORT TO RF IN PORT” will appear.

Step 17. Remove both loads, connect an adapter and/or a cable from the RF Out test port

to the RF In test port and press

ENTER

.

Step 18. Verify that the calibration has been properly performed by checking that the

message “CAL ON” is displayed in the upper left corner of the display.

Setting the Amplitude Scale

The following procedure sets the amplitude scale.

Step 1. Press the

AMPLITUDE

or

LIMIT

key to call up the Scale menu.

Step 2. In SWR mode, press the

TOP

soft key and enter 1, ., 1 using the key pad or the

Up/Down arrow key, and press

ENTER

. (In Return Loss mode, enter 0; in Inser-

tion Gain/Loss mode, enter 100.)

Step 3. In SWR mode, press the

BOTTOM

soft key and enter 1 using the key pad or the

Up/Down arrow key, and press

ENTER

. (In Return Loss mode, enter 54; in In-

sertion Gain/Loss mode, enter –120.)

NOTE:
Figure 4-3
and 4-4 show the typical responses of a valid OSL calibration with a
load attached to the test port with 517 data points.
Figures 4-5,
4-6, 4-7 and 4-8 show typical responses of a valid OSLT calibra-
tion with loads attached to the test port and with the RF Out test port attached
to the RF In test port.
To change the number of data points, press the

SWEEP

key, followed by the

RESOLUTION

soft key. Select 130, 259, or 517 data points.

Saving the Setup

The following procedure can be used to save a setup and calibration in memory.

Step 1. Press the

SAVE SETUP

key.

Step 2. Use the key pad or the Up/Down arrow key to select a memory location (1 - 10),

then press

ENTER

.

Recalling a Setup

The following procedure recalls a setup from memory.

Step 1. Press the

RECALL SETUP

key.