Power On Procedure

The Site Master is designed specifically for field environments and applications requiring
mobility. As such, it is a lightweight, handheld, battery operated unit which can be easily
carried to any location, and is capable of up to 2.5 hours of continuous operation from a
fully charged battery. Built-in energy conservation features allow battery life to be extended
over an eight-hour workday. The Site Master can also be powered by a 12.5 Vdc external
source. The external source can be either the Anritsu AC-DC Adapter (P/N 40-115) or 12.5
Vdc Automotive Cigarette Lighter Adapter (P/N 806-62) provided.

Step 1. Press the

ON/OFF

key.

The Site Master takes about five seconds to perform a series of self-diagnostic
and adjustment routines. At completion, the screen displays the Anritsu logo, the
model number, and the version of firmware installed.

Step 2. Press

ENTER

to continue, or wait for one minute.

The Site Master is now ready for operation.

Calibration Procedure

The following procedure provides the steps necessary to perform Open-Short-Load (OSL)
and Open-Short-Load-Thru (OSLT) calibrations. Calibration is done with the maximum
517 data points and is valid for 130 and 259 data point settings (default is 259). Refer to
Figure 4-1 for setup.

Step 1. Press the

MODE

key.

Step 2. Use the Up/Down arrow key to scroll to

FREQ - SWR

or

RETURN LOSS

for

OSL calibration; or scroll to

INSERTION LOSS

(+6 dBm) or

INSERTION GAIN

(–30 dBm) measurement mode for OSLT calibration..

4-2

Chapter 4 Measurements

HOLD

RUN

START
CAL

AUTO
SCALE

SAVE
SETUP

RECALL
SETUP

LIMIT

MARKER

SAVE
DISPLAY

RECALL
DISPLAY

PRINT

MODE

FREQ/DIST

AMPLITUDE

SWEEP

SYS

ENTER

CLEAR

ESCAPE

ON

OFF

/

1

2

4

5

6

7

8

9

0

3

+

-

.

Site Master S251B

625.0

2500.0

1384.45 MHz

2096.66 MHz

OPEN

LOAD

LOAD

SHORT

CALIBRATION

RF OUT TEST PORT

RF IN TEST PORT

TEST PORT CABLE (OPTIONAL)

TEST PORT CABLE (OPTIONAL)

Figure 4-1.

One-port Measurement Calibration Setup