Making power mea sure ments 4-27, En tering power mon i tor mode 4-27, Zeroing the power mon i tor 4-27 – Anritsu S251B User Manual

Page 73: Mea suring high in put power levels 4-27, Dis playing power in dbm and watts 4-27, Dis playing rel a tive power 4-27, Making power measurements -27, Entering power monitor mode -27, Zeroing the power monitor -27, Measuring high input power levels -27

Advertising
background image

Making Power Measurements

Power measurement is accomplished using a broadband (1 MHz to 3000 MHz) RF detector,
P/N 5400-71N50. The power monitor displays measured power in dBm or Watts.

Entering Power Monitor Mode

Step 1. Press the

MODE

key.

Step 2. Use the Up/Down arrow key to scroll to

POWER MONITOR

. Press

ENTER

to

select power monitor mode.

Zeroing the Power Monitor

Step 3. With no power applied to the DUT, press the

ZERO

soft key from the Power

menu. Wait a few seconds while the Site Master accumulates samples of the qui-
escent power level. When complete,

ZERO ADJ: ON

is displayed in the message

area.

Measuring High Input Power Levels

Step 4. Insert an attenuator between the DUT and the RF detector, sufficient to insure

that the input power to the Site Master is no greater the 20 dBm.

Step 5. Press the

OFFSET

soft key.

Step 6. Enter the attenuation in dB using the keypad. Press

ENTER

to complete the en-

try. The message area will show

OFFSET is ON

along with the entered value in

dB.

Displaying Power in dBm and Watts

Step 7. Press the

UNITS

soft key to display power in Watts.

Displaying Relative Power

Step 8. With the desired base power level input to the Site Master, press the

REL

soft

key. The message area will show

REL: ON

and the power reading will indicate

100%.

Step 9. Press the

UNITS

soft key to display power in dBm. Since REL is ON, the power

reading will be in dBr, relative to the base power level.

4-27/4-28

Chapter 4 Measurements

Advertising