Step 24. Enter 6, 2, 5 using the key pad or the Up/Down arrow key. Press

ENTER

to set

the Start Frequency to 625 MHz.

Step 25. Use the Up/Down arrow key to scroll to the

START FREQ LIMIT

parameter and

press

ENTER

to edit the value.

Step 26. For FREQ-SWR, enter 1, ., 1, 2. For FREQ-RETURN LOSS, enter 2, 5. Press

ENTER

to set the Start Frequency Limit value.

Step 27. Use the Up/Down arrow key to scroll to the

STOP FREQ

parameter and press

ENTER

to edit the value.

Step 28. Enter 2, 5, 0, 0 using the key pad or the Up/Down arrow key. Press

ENTER

to

set the Stop Frequency to 2500 MHz.

Step 29. Use the Up/Down arrow key to scroll to the

STOP FREQ LIMIT

parameter and

press

ENTER

to edit the value.

Step 30. For FREQ-SWR, enter 1, ., 2. For FREQ-RETURN LOSS, enter 2, 1. Press

ENTER

to set the Stop Frequency Limit value.

Step 31. Press

ENTER

at OK to accept the values and activate the Segment 1 limit line.

NOTE:
Alternatively, if the Start and Stop frequencies are the full span of the Site Mas-
ter, simply edit the Start and Stop frequency Limits and select the

FULL SPAN

soft key. Press

ENTER

at OK to accept and activate the Segment 1 limit line.

4-12

Chapter 4 Measurements