Ca ble loss mea sure ment 4-14, Set ting the am pli tude scale 4-14, Set ting the seg ment limit lines 4-14 – Anritsu S251B User Manual

Page 60: Cable loss measurement -14, Setting the amplitude scale -14, Setting the segment limit lines -14, Cable loss measurement

Advertising
background image

Cable Loss Measurement

The following frequency domain measurement is an example of a Cable Loss measurement
made over a selectable frequency range following the Marker setting steps 32 to 35.

NOTE:
If SWR/Return Loss measurement has not been done, calibration must be per-
formed before determining the cable loss.

Step 36. Disconnect the load from the end of the cable and replace it with a short.

Step 37. Press the

MODE

key.

Step 38. Use the Up/Down arrow key to scroll to

CABLE LOSS - ONE PORT

.

Step 39. Press

ENTER

to select Cable Loss measurement mode.

Setting the Amplitude Scale

The following procedure sets the scale display.

Step 40. Press the

AMPLITUDE

key to call up the Scale Menu.

Step 41. Press the TOP soft key and enter 0 using the keypad or Up/Down arrow key.

Press

ENTER

to set the top scale.

Step 42. Press the BOTTOM soft key and enter 1 using the keypad or Up/Down arrow

key. Press

ENTER

to set the bottom scale.

NOTE:
An alternative method of setting the amplitude scale is to use the

AUTO SCAL

E

key.

Setting the Segment Limit LInes

The following procedure sets the segment limit lines.

Step 43. Press the

AMPLITUDE

or

LIMIT

key to call up the Segment Limit Line menu.

Step 44. Press the

LIMIT

soft key to activate the segment limit line table.

Step 45. Use the Up/Down arrow key to scroll to segment #1 and press the

EDIT SEG-

MENT

soft key to change the parameter values for segment #1.

Step 46. Use the Up/Down arrow key to scroll to the

START FREQ

parameter and press

ENTER

to edit the value.

Step 47. Enter 8, 2, 4 using the key pad or the Up/Down arrow key. Press

ENTER

to set

the Start Frequency to 824 MHz.

Step 48. Use the Up/Down arrow key to scroll to the

START FREQ LIMIT

parameter and

press

ENTER

to edit the value.

Step 49. Enter ., 5. Press

ENTER

to set the Start Frequency Limit value.

Step 50. Use the Up/Down arrow key to scroll to the

STOP FREQ

parameter and press

ENTER

to edit the value.

4-14