DTF Measurement (Determining the quality of the cable.)

NOTE:
If cable length measurement has not been done, calibration must be done be-
fore determining the quality of the cable (steps 1-15).

Step 24. Disconnect the Short and connect the Load to the test port extension cable

Setting the Amplitude Scale

Step 25. Press the

AMPLITUDE

or

LIMIT

key to call up the Scale Menu..

Optional: Press the

AUTO SCALE

key to optimize the amplitude scale.

Step 26. In SWR mode, press the

TOP

soft key and enter 1, ., 0, 5 using the keypad or

Up/Down arrow key. Press

ENTER

to set the top scale. (In Return Loss mode,

enter 20.)

Step 27. In SWR mode, press the

BOTTOM

soft key and enter 1 using the keypad or

Up/Down arrow key. Press

ENTER

to set the bottom scale. (In Return Loss

mode, enter 54.)

Setting the Markers

Step 28. Press the

MARKER

key to call up the Markers menu.

Step 29. Press the

M2

soft key to select the M2 marker function. (Note: pressing the

ON/OFF

soft key activates or deactivates the M2 marker function.)

Step 30. Press the

MARKER TO PEAK

soft key to find the maximum SWR or Return

Loss reading. (See Figures 4-15 and 4-16.)

NOTE:
Refer to page 4-4
for instructions on saving and recalling a setup or display.