Symbols

Table 3-1 provides a listing of the symbols used as condition indicators on the LCD display.

Self Test

At turn-on, the Site Master runs through a series of quick checks to ensure the system is
functioning properly. Note that the battery voltage and temperature are displayed in the
lower left corner below the self test message. If the battery is low, or if the ambient temper-
ature is not within the specified operational range, Self Test will fail. If Self Test fails and
the battery is fully charged and the Site Master is within the specified operating range, call
your Anritsu Service Center.

3-23

Chapter 3 Functions and Operations

Icon

Symbol

Site Master is in Hold for power conservation. To resume sweeping, press
the

RUN/HOLD

key. After 10 minutes without a key press, the Site Master

will automatically activate the power conservation mode.

Integrator Failure

Lock fail indication. Check battery. (If the Site Master fails to lock with a
fully charged battery, call your Anritsu Service Center.)

When calibration is performed, the Site Master stores the ambient temper-
ature. If the temperature drifts outside the specified range, this indicator
will flash. A recalibration at the current temperature is recommended.

Indicates the remaining charge on the battery. The inner white rectangle
grows longer as the battery charge depletes.

Indicates internal data processing.

Table 3-1.

LCD Icon Symbols