Se lecting the fre quency range 4-3, Per forming a cal i bra tion 4-3, Selecting the frequency range -3 – Anritsu S251B User Manual

Page 49: Performing a calibration -3, Selecting the frequency range, Performing a calibration, Step 3. press the, Key to select the measurement mode, Key. step 5. press the, Chapter 4 measurements

Advertising
background image

Step 3. Press the

ENTER

key to select the measurement mode.

Selecting the Frequency Range

The following procedure selects the frequency range for the calibration.

Step 4. Press the

FREQ/DIST

key.

Step 5. Press the

F1

soft key.

Step 6. Enter 6, 2, 5 or a user-defined frequency using the key pad or the Up/Down ar-

row key.

Step 7. Press

ENTER

to set F1 to 625 MHz or the user-defined frequency.

Step 8. Press the

F2

soft key.

Step 9. Enter 2, 5, 0, 0 or a user-defined frequency using the keypad or the Up/Down ar-

row key.

Step 10. Press

ENTER

to set F2 to 2500 MHz or the user-defined frequency.

Step 11. Check that the FREQ (MHz) scale in the display area indicates the new fre-

quency start and stop values (for example: F1 = 625 MHz, F2 = 2500 MHz).

Performing a Calibration

The following procedures detail how to perform the calibrations. Steps 12 through 15 de-
scribe OSL calibration, while steps 16 and 17 apply only to OSLT calibration.

Step 12. Press the

START CAL

key. The message “CONNECT OPEN TO RF OUT

PORT” will appear in the display.

Step 13. Connect the calibrated Open and press the

ENTER

key. The messages “Mea-

suring OPEN” and “CONNECT SHORT TO RF OUT” will appear.

Step 14. Remove the Open, connect the calibrated Short and press the

ENTER

key. The

messages “MEASURING SHORT” and “CONNECT LOAD TO RF OUT” will
appear.

4-3

Chapter 4 Measurements

HOLD

RUN

START
CAL

AUTO
SCALE

SAVE
SETUP

RECALL
SETUP

LIMIT

MARKER

SAVE
DISPLAY

RECALL
DISPLAY

PRINT

MODE

FREQ/DIST

AMPLITUDE

SWEEP

SYS

ENTER

CLEAR

ESCAPE

ON

OFF

/

1

2

4

5

6

7

8

9

0

3

+

-

.

Site Master S251B

625.0

2500.0

1384.45 MHz

2096.66 MHz

LOAD

OPEN

LOAD

SHORT

RF OUT TEST PORT

RF IN TEST PORT

TEST PORT CABLE
(OPTIONAL)

TEST PORT CABLE

(OPTIONAL)

Figure 4-2.

Two-Port Measurement Calibration Setup