Rockwell Automation 1734-VHSC24 Very High-Speed Counter Modules User Manual User Manual

Page 34

Advertising
background image

Publication 1734-UM003B-EN-P - August 2000

3-4 Very High-Speed Counter Module Input and Output Data

EEPROM Fault status bit (EF) - If a fault is detected with the
EEPROM during power up tests, this bit is asserted to 1. It indicates
that the content of the EEPROM has been corrupted, most likely
caused by loss of power during an executing write.

Not Ready status bit (NR) - Whenever power is applied to the
module, the hardware must be initialized. During this time, the NR bit
is asserted and the green module status indicator flashes.

Output Fault Status indicators (FS) - where bit 11 is output 1 and
bit 10 is output 0. A 1 indicates the output is either shorted or open.

Output Status indicators (OS) - where bit 9 is output 1 and bit 8 is
output 0. A 1 indicates the output is ON, 0 it is OFF.

Z input Status (ZS) - This bit indicates the present status of the Z
input (1 indicates Z is ON, 0 indicates Z is OFF). This bit is unaffected
by Z Invert, ZI, in the Counter Configuration word.

B input Status (BS) - This bit indicates the present status of the B
input (1 indicates B is ON, 0 indicates B is OFF).

A input Status (AS) - This bit indicates the present status of the A
input (1 indicates A is ON, 0 indicates A is OFF).

C[1,0] Stored data count - This count cycles through [ 0 0 ], [ 0 1 ], [ 1
0 ], [ 1 1 ], [ 0 0 ]

… Each time the stored/accumulated count words are

updated, C[1,0] is incremented. This feature assumes the host’s sample
rate (including network delay and program scan) is as fast or faster
than the frequency of the event which updates C[1,0].

Zero frequency Detected (ZD) - This bit is operational when
frequency configurations are programmed (configurations: period/rate
[5], continuous/rate [6], rate measurement [7]).

In period/rate [5] and continuous/rate [6] configurations, counts
are acquired during the ON state of the Z input. At very low
frequencies the counter saturates, indicating a zero frequency
detect. The time it takes to determine a zero frequency in these
two configurations can be as long as 6.7 seconds ( 16,777,216
counts x 1/5 MHz x 2 half cycles of Z ).

In rate measurement [7] configuration pulses on the A input are
counted over a sample interval specified by the time base. The
time it takes to determine a zero frequency in this configuration
is determined by the sample interval (for example, time base =
0.300 second therefore 300 milliseconds to determine ZF).

Advertising
This manual is related to the following products: