9.13 Electromagnetic compatibility (EMC)

■■■■■■■■■■■■■■■■■■■■■■

90

■■■■■■■■

894 Professional CVS

9.13

Electromagnetic compatibility (EMC)

Emission

Standards ful-
filled

EN/IEC 61326-1

EN/IEC 61000-6-3

EN 55011 / CISPR 11

EN/IEC 61000-3-2

EN/IEC 61000-3-3

Immunity

Standards ful-
filled

EN/IEC 61326-1

EN/IEC 61000-6-2

EN/IEC 61000-4-2

EN/IEC 61000-4-3

EN/IEC 61000-4-4

EN/IEC 61000-4-5

EN/IEC 61000-4-6

EN/IEC 61000-4-8

EN/IEC 61000-4-11

EN/IEC 61000-4-14

EN/IEC 61000-4-28

9.14

Ambient temperature

Nominal function
range

0 - +45 °C
(at a maximum of 85% relative humidity)

Automatic interior
temperature mon-
itoring

> 70 °C

Storage

–40 - +70 °C

Transport

–40 - +70 °C