9.7 Resolution
■■■■■■■■■■■■■■■■■■■■■■
88
■■■■■■■■
894 Professional CVS
9.7
Resolution
Potential applied
15 µV
Potential mea-
sured
150 µV
Current applied
0.0031% of the current measuring range
Current measured
0.0031% of the current measuring range
Current measured
in the smallest
current measuring
range (63 pA)
2 fA
9.8
Calibrator
Reference voltage
2.5 V
R17.8
17.8
Ω
R178
178
Ω
R1.9k
1,900
Ω
R18k
18,000
Ω
R180k
180,000
Ω
R1.8M
1,800,000
Ω
R18M
18,000,000
Ω
Validity of the cer-
tification
Two years after activation
9.9
Stirrer
Stirring rate
200 - 3,000 min
-1
Stability
±2%