Metrohm 894 Professional CVS User Manual
Page 64
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
894 Professional CVS
5 Insert the measuring vessel into the holder (1-3).
6
CAUTION
If the measuring head arm is lowered carelessly, this may result in
injuries to the hand.
Make sure that you do not trap a finger between the measuring
head arm and the instrument housing.
Lower the measuring head arm.
7 Place the drip pan (6.2711.090) into the drip pan holder (1-8).
8 Connect the devices electrically (see Chapter 3.4, page 38).
Activating the calibrator
The calibrator built into the 894 Professional CVS guarantees a very high
measuring accuracy for two years after the initial start-up. You can find
more detailed information about the calibrator in Chapter 6.7, page 78.
In viva, proceed as follows:
1 Select the 894 Professional CVS in the device table of the Configu-
ration section of the program.
2 In the device table, click on the Edit button and select Properties....
3 Open the Calibrator tab.