6 Operation and maintenance

894 Professional CVS

■■■■■■■■

77

7 Tighten the screw on the right-hand side of the measuring head con-

nector plate again.

6.6

Adjusting the PEEK needle in the Sample Processor

If the entire sample volume is to be transferred from the Sample Pro-
cessor, then the work position on the tower of the Sample Processor has
to be set in such a way that the PEEK needle is located no more than
0.5 mm above the base of the sample vial. The work position settings
must be defined in viva.

If required, adjust the PEEK needle in viva as shown in the figure:

max. 0.5 mm

Figure 49

Adjusting the PEEK needle in the Sample Processor