3 connecting a sample processor, Connecting a sample processor – Metrohm 894 Professional CVS User Manual

Page 58

Advertising
background image

3.4 Connecting devices electrically

■■■■■■■■■■■■■■■■■■■■■■

48

■■■■■■■■

894 Professional CVS

NOTE

Under Tubing parameters, adapt the parameters Length and
Diameter to the actual installation. This ensures that functions
such as Prepare or Empty work properly in viva. The lengths and
diameters of the tubing connections incorporated in the measur-
ing head arm can be found in Chapter 8.1, page 85.

5 Click on Yes if you are using the already configured dosing unit.

The dosing unit will be automatically displayed in the Dosing units
subwindow of the Configuration section of the program.

NOTE

Check the Tubing parameters and adapt them to the actual
installation, if necessary.

3.4.3

Connecting a Sample Processor

The following Sample Processors can be used in conjunction with the 894
Professional CVS:

858 Professional Sample Processor

919 IC Autosampler plus

814 USB Sample Processor

815 Robotic USB Sample Processor XL

Sample Processors are connected to the power supply with a power sup-
ply cable and and to the PC with a controller cable.

WARNING

An incorrect supply voltage can damage the instrument. Operate the
instrument only with the supply voltage specified for it.

Connecting to power supply and PC

Proceed as follows:

1 Connect the Sample Processor to the power supply with the power

supply cable (6.2122.0x0).